Problem |
Possible Causes |
Solutions |
Light color, low sensitivity |
1. Long transportation time with high temperature. |
Shorter transportation time, pack the kit with foam box and ice bags. |
2. Without fully equilibration of the kit to room temperature before use. |
Bring all reagents to room temperature (25°C) before use. |
|
3. Incubator temperature less than 37°C. |
Make sure the steady Incubator temperature. |
|
4. Insufficient incubation time. |
Accurate timing. |
|
5. Incorrect washing process. |
See washing procedure in the manual, The wash procedure is critical. |
|
6. Insufficient fluid absorption of pipette, too much water or uncleanness inside of suction nozzle. |
Adjust pipette, use clean suction nozzle matching with pipette tightly, dispose suction nozzle after use. |
|
7. Distilled Water contamination. |
Use fresh qualified distilled water. |
|
8. Insufficient substrate working time. |
Accurate timing. |
|
9. low concentration in HRP or detection antibody. |
Raise content of Avidin-HRP and detection antibody. |
|
High background |
1. Insufficient washing. |
Prepare Wash Concentrate accurately. See washing procedure in the manual, The wash procedure is critical. |
2. Sample contamination. |
Collect fresh sample or store at low temperature to avoid pollution. |
|
3. Incubator temperature higher than 37°C or too long reaction time. |
Adjust Incubator temperature, Accurate timing. |
|
4. Reuse suction nozzle, Insufficient washing or disinfecting. |
Dispose suction nozzle after use. |
|
5. Distilled Water contamination. |
Use fresh distilled water. |
|
6. Mix reagents like enzyme. |
Do not mix reagents from different batches. |
|
7. Too many samples cause long addition time and reaction time. |
Control of reaction time to avoid adding too many samples once. |
|
8. Insufficient sealing cause unspecific adsorption. |
Extend sealing time. |
|
9. High concentration in HRP or detection antibody, Insufficient washing. |
Reduce content of Avidin-HRP and detection antibody. |
|
Poor reproducibility |
1. Variations in samples quantity and addition time. |
Controlling of addition time. |
2. Variations in Incubation time, washing methods and operators. |
Make the same experiment condition. |
|
3. Variations in Sample addition. |
Samples should be mix well before diluting, use the same transferpettor and set suction nozzle tightly. |
|
Whiteboard |
1. Substrate deteriorated. |
Replace new substrate. |
2. Wash Solution error. |
See Dilution ratio in the manual. |
|
3. Miss Enzyme conjugate. |
Do not miss Enzyme conjugate. |
|
4. Mistake stop solution for wash buffer or substrate. |
Check the labels of each reagent carefully. |
|
5. Process sequence error. |
Make sure the correct process sequence in the manual. |